3D scanning is faster than plaster casting but requires the same care. A poor scan produces a model that does not match reality – and this may not be discovered until the aid is to be tested.

  • 01Calibrate the scanner according to the manufacturer's instructions before you start the day
  • 02Ensure even, glare-free light around the body part
  • 03Place the user in the correct load or rest position
  • 04Ask the user to keep still during scanning – movement causes distortion
  • 05Scan from multiple angles with complex geometry eg a foot with high arch
  • 06Review the 3D model for holes or noise before closing the session
  • 07Name and save the file with user ID date and body part according to the clinic's system
  • 08Back up the file before transferring it to the CAD program
Error sourceConsequenceHow to avoid it
Movement during scanningDistorted modelUse support or holding point for the user
Uneven light.Holes in scanning dataUse even lighting and avoid shadows
Incorrect loadingIncorrect shape of foot or legScan in the position the aid will be used in